Noniterative complex permittivity retrieval using calibration-independent waveguide measurements

MakaleYalnız künyeScopus SJRQ1

Özet

Calibration-independent nonresonant measurements can be employed to eliminate the need for calibration before microwave measurements. These methods generally assume that the location of the sample in its measurement cell is known after the sample is positioned or shifted within the cell and that the length of the measurement cell is known. In addition, in general these methods necessitate some sort of numerical analysis to retrieve the complex permittivity of the sample. In this research paper, we propose a calibration-independent nonresonant waveguide method to first eliminate measurement errors arising from inaccurate knowledge of the location of the sample within its cell (after shifting) and arising from improper knowledge of the length of the measurement cell and second to determine the complex permittivity explicitly. In addition, our proposed method needs scattering parameter measurements from three different measurement configurations – another improvement with respect to a similar calibration-independent method requiring measurements from four different configurations. Furthermore, it does not necessitate that the sample be flushed to the measurement cell. We analyzed repeatability of cell connection and investigated uncertainty in the measurement of complex permittivity for a change in length of the sample and the cell and when air gap present between sample surfaces and guide walls. We validated the proposed method by measuring the complex permittivity of polyethylene sample and distilled water at X-band (8.2–12.4 GHz).

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/bayburt/13603Resmi atıflarda kullanılan kalıcı adres
DOI
https://doi.org/10.1016/j.sna.2017.07.018Yayıncı resmi DOI adresi
İndeks kayıtları
Kaydın indeks veri tabanlarındaki sayfası (aktarımda alınan kayıt numarasıyla). Erişim, indeksin kendi koşullarına bağlıdır.
Atıflar
Scopus
8
WoS
8
OpenAlex
8
Her kaynağın kendi saydığı toplam atıf; aktarım ve zenginleştirme sırasında alındı.
Tam Kayıt
AlanDeğerDil
dc.contributor.authorHasar, Ugur Cem
dc.contributor.authorKaya, Yunus
dc.date.accessioned2026-10-29T19:11:12.677892130Z
dc.date.available2026-10-29T19:11:12.677892130Z
dc.date.issued2017
dc.description.abstractCalibration-independent nonresonant measurements can be employed to eliminate the need for calibration before microwave measurements. These methods generally assume that the location of the sample in its measurement cell is known after the sample is positioned or shifted within the cell and that the length of the measurement cell is known. In addition, in general these methods necessitate some sort of numerical analysis to retrieve the complex permittivity of the sample. In this research paper, we propose a calibration-independent nonresonant waveguide method to first eliminate measurement errors arising from inaccurate knowledge of the location of the sample within its cell (after shifting) and arising from improper knowledge of the length of the measurement cell and second to determine the complex permittivity explicitly. In addition, our proposed method needs scattering parameter measurements from three different measurement configurations – another improvement with respect to a similar calibration-independent method requiring measurements from four different configurations. Furthermore, it does not necessitate that the sample be flushed to the measurement cell. We analyzed repeatability of cell connection and investigated uncertainty in the measurement of complex permittivity for a change in length of the sample and the cell and when air gap present between sample surfaces and guide walls. We validated the proposed method by measuring the complex permittivity of polyethylene sample and distilled water at X-band (8.2–12.4 GHz).
dc.description.sponsorshipScience and Technological Research Council of Turkey (TUBITAK) [114E495]
dc.identifier.doi10.1016/j.sna.2017.07.018
dc.identifier.endpage666
dc.identifier.issn0924-4247
dc.identifier.scopus2-s2.0-85026487591
dc.identifier.startpage654
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/bayburt/13603
dc.identifier.volume263
dc.identifier.wosWOS:000409149900077
dc.language.isoen
dc.publisherElsevier B.V.
dc.relation.ispartofSensors and Actuators, A: Physical
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectCalibration-independent
dc.subjectComplex permittivity
dc.subjectExplicit determination
dc.subjectSample shifting
dc.subjectWaveguide measurements
dc.subject.classificationEngineering, Electrical & Electronic
dc.subject.classificationInstruments & Instrumentation
dc.subject.classificationEngineering
dc.subject.otherCalibration
dc.subject.otherCells
dc.subject.otherCytology
dc.subject.otherNumerical methods
dc.subject.otherPermittivity
dc.subject.otherWaveguides
dc.subject.otherComplex permittivity
dc.subject.otherDistilled water
dc.subject.otherExplicit determination
dc.subject.otherMeasurement cell
dc.subject.otherMeasurement configuration
dc.subject.otherMeasurement of complex permittivity
dc.subject.otherScattering parameter measurement
dc.subject.otherWaveguide measurements
dc.subject.otherUncertainty analysis